Product Details:
|
Deceleration: | 10 Deceleration Profiles | Acceleration: | 10 Acceleration Profiles |
---|---|---|---|
Maximum Speed: | 4500 Rpm | Power Requirements: | 110V/220-240V, 50/60Hz |
Certifications: | CE, ISO, IUV Certified | Rotor Compatibility: | Red And White Blood Cell Rotors |
Capacity: | 12 Tubes | ||
Highlight: | Cell Cleaning Washer Centrifuge,Red blood Cell Cleaning Washer Centrifuge,HLA Examinations Washer Centrifuge |
The K450 Cell Washing Centrifuge is designed for lymphocyte cleaning in HLA examinations and red blood cell cleaning for freezing experiments in laboratory settings. Ideal for laboratories conducting HLA examinations and freezing experiments, this centrifuge ensures precise and effective cell washing, which is essential for accurate outcomes. Its multifunctional design caters to diverse applications in the medical and biochemical domains, making it a versatile choice for laboratories requiring efficient cell-washing capabilities.
Robust Construction: High-quality steel structure with built-in blast-proof steel chamber and autoclave-able 304# stainless steel chamber for resistance against detergents and disinfection liquids.
Efficient Drive System: Frequency conversion motor drive with microprocessor control, utilizing a flexible shaft drive system for smooth operation, minimal vibration, and low noise.
Convenient Monitoring: LED screen displays Speed, Timer, RCF, Rotors, Program(Memory), Acceleration& Deceleration information, and Error codes.
Program Flexibility: Pre-programmed settings with memory for lymphocyte cleaning and red blood cell separation. Ten speed stages with rapid acceleration (7 seconds) and deceleration. Fast shutdown in just 5 seconds.
Seamless Operation: Equipped with double-layer silicone seal rings for effective sealing. Automatic lid-locking and holding during rotor operation. Easily exchangeable rotors with a special connector for quick loading and unloading.
Diagnostic System: Self-diagnostic system integrated with high-strength chamber-coating and protective features. Automatic rpm/rcf conversion for enhanced convenience.
Effortless Operation: Automatic lid-locking and holding during rotor run. Easily interchangeable rotors with a special connector for swift loading and unloading.
Diagnostic Intelligence: Self-diagnostic system with robust chamber-coating and protective measures. Automatic rpm/rcf conversion simplifies operation.
Operational Safety: Automatic lid-locking and holding during rotor run. Special rotor connector ensures secure loading and unloading.
Protective Coating: Self-diagnostic system includes a high-strength chamber-coating and protective features for enhanced safety. Automatic rpm/rcf conversion enhances operational safety.
Technical Specifications:
Speed range(rpm) | 0-4500rpm adjustable | Max Capacity(ml) | 12x15ml |
RCF range (xg) | 0-2000xg adjustable | ACC/DEC | 1-10 rates |
Motor | Inverter motor direct drive | Dimension(mm) | 430*320*220mm |
Noise(dBA) | ≦55dB(A) | Net weight(without rotor) | 18KG |
Power supply | AC220V,50HZ,5A,550W |
Matched Rotors
For lymph ball cleaning | ||||||
Rotor Name | Program | Capacity | Speed r/min | RCF xg |
Centrifugal time (variable range) |
Applications |
White blood cell washing rotor | P1 | 12×2ml | 4500 | 2000 |
180seconds (30~300seconds) |
Isolation of lymphocytes, isolation of cultured cells |
P2 | 12×1.5ml | 3200 | 1000 | 3 seconds | Platelet removal | |
P3 | 12×0.5ml | 3200 | 1000 |
60 seconds (30~120seconds) |
Cleaning of the lymphatic bulb | |
Red blood cell washing rotor | ||||||
Rotor Name | Program | Capacity | Speed r/min | RCF xg |
Centrifugal time (variable range) |
Applications |
Red blood cell washing rotor
|
P4 | 12×2ml | 2100 | 500 |
60 seconds (30~120seconds) |
Blood group determination, blood clotting reaction observation |
P5 | 12×2ml | 3200 | 1000 |
15 seconds (5~15seconds) |
Cross-fit test, antiglobulin test | |
P6 | 12×2ml | 3200 | 1000 |
60seconds (30~120seconds) |
Washing of blood cells, extraction of serum and plasma from dissociated solutions | |
Angle Rotor
|
P7 | 12×15-10-5ml | 3200 | 1247 | 60 seconds (5~15 seconds) | Cross-matching blood test |
WHY CHOOSE US
Our Factory Introduction
We boast a standardized production workshop operated by a dedicated team committed to upholding seriousness and responsibility in every product manufactured. Our R&D division comprises versatile professionals including design engineers, technology specialists, qualified experts, and process control managers. With a stringent adherence to our quality process control system, our qualified specialists rigorously oversee the entire production process, encompassing product design, research and development, testing, production, quality inspection, and packaging.